4Department of Radiation Oncology, Apollo Multispeciality Hospital, Kolkata-India DOI : 10.5505/tjo.2022.3783 OBJECTIVE
This study assesses the use of the flattening filter free (FFF) photon beam in hybrid treatment planning for breast carcinoma and the various dosimetric indices of planning target volume (PTV) and organs at risk (OARs).
METHODS
We selected 15 female breast cancer patients treated with Field-in-Field technique to a dose of 40 Gy/15
fractions. Retrospectively, hybrid intensity-modulated radiation therapy (IMRT) plans with FF, and FFF
photon beams were created. To make the hybrid IMRT plan, a planning weightage of 60% (3DCRT):
40% (IMRT) was applied. The dose-volume histograms (DVH) were assessed for various dosimetric
indicators for PTV and OARs.
RESULTS
A UDI scoring of 1.090±0.023 and 1.078±0.024 was observed between 6FF_Hybrid and 6FFF_Hybrid treatment
plans with p<0.05. ID for I/L lung was 6387±1658.51 (Gy-L) and 6347.056±1643.41 (Gy-L) in 6FF_
Hybrid and 6FFF_Hybrid IMRT plan (p=0.05), respectively. ID value to heart was 2272.52±1086.63 (Gy-
L) and 2212.40±1059.49 (Gy-L) in 6FF_Hybrid and 6FFF_Hybrid IMRT plans (p>0.05), respectively. The
beam on time (BOT) values of 0.882±0.08 (6FF_Hybrid) and 0.5436±0.07 (6FFF_Hybrid) were reported.
CONCLUSION
Hybrid IMRT planning with an FFF photon beam offers comparable target coverage, conformality, and
homogeneity with greater OAR sparing and faster treatment time.