The end-to-end (E2E) testing method enables understanding the difficulties and uncertainties in treating any specific case type. This study was focused on bilateral metallic implant cases.
METHODS
The study was performed on a cylindrical phantom of Perspex with holes for implant inserts. Two stainless
steel metal rods of 7.5-8.0 g/cc mass density were inserted in the phantom. The ionization chamber
CC13 was kept at a 5 cm depth in the phantom. The phantom was scanned on a computed tomography
simulator in pelvis protocol with a 1mm slice thickness. The scans were imported to the contouring
station without applying artifacts correction. Chamber volume was contoured as gross tumor volume
(GTV); margin to GTV, clinical target volume, and planning target volume were created. Four isocentric
plans (Conventional, three-dimensional conformal radiotherapy[3D-CRT], intensity-modulated
radiotherapy [IMRT], and volumetric-modulated radiotherapy [VMAT]) were generated for two LinacsTruebeam
(TB)-sTx and 2300-CD. The conventional plan was a single anterior field, 3D-CRT was four
field box techniques, IMRT was seven field plan, and VMAT was with two complete arc. Pre-treatment
verification was done using CBCT. Four plans were created on helical tomotherapy with different prescriptions
and delivered using MVCT guidance.
RESULTS
In conventional plans, variations were -1.40%, -1.57%, and for 3DCRT, variations were -5.08% and
-4.93%, for IMRT, the differences between measured and TPS doses were 1.84 % and -1.55% for VMAT
plans, and the variations were 0.68% and -0.88% for TB and 2300-CD, respectively. The tomotherapy
plans with gradient showed deviations more significant than 3%. Similarly, the variations for single prescription
plans were within 3%.
CONCLUSION
The phantom design used in the test provided a comprehensive understanding of simulation and delivery
problems.