TURKISH JOURNAL OF ONCOLOGY 2009 , Vol 24 , Num 1
Evaluation of spinal cord doses in patients with head and neck cancer receiving radiotherapy
Vildan KAYA, Melek Gamze AKSU, Aylin Fidan KORCUM, Nina TUNÇEL
Akdeniz Üniversitesi Tıp Fakültesi, Radyasyon Onkolojisi Anabilim Dalı OBJECTIVES
Spinal cord doses were analyzed for head and neck cancer patients receiving radiotherapy with three-field technique.

METHODS Eighteen patients (12 males, 6 females) with head and neck carcinoma were evaluated. Spinal cord doses were evaluated for each treatment phase. Biologically effective doses (BED) were calculated for the maximum spinal doses in all patients and compared to the 92 Gy, which was the BED of the spinal cord tolerance dose of 46 Gy.

RESULTS
The maximum spinal cord dose ranged from 4399 to 4954 cGy. The median dose at the junction of the lateral and anterior fields was 4131 cGy. BED of the maximum spinal doses ranged between 75-98 Gy (median 91), and in seven patients it was over 92 Gy; total spinal dose was over 46 Gy in 13 patients.

CONCLUSION
Spinal cord doses in each phase should be calculated separately with BED values for assessment of the cumulative spinal cord doses. Keywords : Head and neck carcinoma, spinal cord, radiotherapy; radiotherapy dosage