TURKISH JOURNAL OF ONCOLOGY 2010 , Vol 25 , Num 4
Verification of cone-beam CT functions for Varian Clinac® IX linear accelerator
Songül ÇAVDAR KARAÇAM, Başak İKTUEREN, Ezgi IŞIKTAŞ, Sedat KOCA, Gülyüz ATKOVAR
İstanbul Üniversitesi Cerrahpaşa Tıp Fakültesi, Radyasyon Onkolojisi Anabilim Dalı, İstanbul OBJECTIVES
The On-Board Imager® (OBI) system is designed to correct for motion and setup errors in patients undergoing radiation therapy. The OBI system provides three imaging modes: radiographic and fluoroscopic image acquisition and three-dimensional cone-beam computed tomography (CBCT) acquisition. The use of this new technology necessitates a comprehensive quality assurance (QA) program. In this study, we describe the procedures of test items included in the QA program and present the results of measurements.

METHODS
We used Catphan® 600 phantom for CBCT image quality tests. We verified density resolution, spatial linearity measurement, image uniformity measurements, and high resolution and low contrast resolution.

RESULTS
All measured values were in accordance with acceptance and limit values.

CONCLUSION
Use of the tests over extended periods showed that the OBI system has stable image quality. While the results of the measurements were not exactly the same, they were observed to be in acceptable limits. Keywords : Quality control; cone-beam CT; verification