TURKISH JOURNAL OF ONCOLOGY
2010 , Vol 25 , Num 4
Verification of plaque simulator dose distributions using GAFCHROMIC® EBT film
İstanbul Üniversitesi Onkoloji Enstitüsü, Klinik Onkoloji Anabilim Dalı, Radyasyon Onkolojisi Bilim Dalı, İstanbul
OBJECTIVESOff-axis dose profile and central-axis dose distributions were measured using GAFCHROMIC® EBT film and calculated using BEBIG plaque simulator for COMS plaques of 14, 16, 18 and 20 mm diameter fully loaded with I-125 seeds. Results were compared to verify dose distribution as calculated by BEBIG.
METHODS
Measurements were made on polystyrene eye phantom placed
in a wax head phantom.
RESULTS
The results of GAFCHROMIC® EBT film were 13% lower
than the results using BEBIG plaque simulator. This can be
attributed to the facts that the model eye does not reflect the
human eye exactly and that our measurement devices are inadequate.
CONCLUSION
The BEBIG treatment planning computer can be used considering
the shortcomings with respect to the first few mm and remote
off-axis points. Furthermore, dose verification in patients
can be made using GAFCHROMIC® EBT film on a model eye
phantom.