TURKISH JOURNAL OF ONCOLOGY 2013 , Vol 28 , Num 1
Repositining of prostate cancer patients, using implanted gold seeds into prostate and electronic portal imaging device
Nedret COŞKUN,2 Nezahat OLACAK,1 Sinan HOCA,1 Serdar ÖZKÖK,1 Arif Bülent ARAS1
1Ege Üniversitesi Tıp Fakültesi Hastanesi, Radyasyon Onkolojisi Anabilim Dalı, İzmir
2Manisa Devlet Hastanesi, Radyasyon Onkolojisi Anabilim Dalı, Manisa
OBJECTIVES
To investigate the feasibility of target focused treatment and it's advantages on standard treatment position verifying tecniques using gold markers and electronic portal imaging device (EPID).

METHODS
Marker placed 10 patient's films were taken, transferred to planning system and distance between markers was measured. Port images were taken by EPID and Template Matching option was used for matching ports with digitally reconstructed radiograph. The distance between markers and the changes were determined with analytic process. Systematical and random set-up errors were determined.

RESULTS
No important changes on marker positions and no radiation toxicity was observed. Verifying the treatment positions according to markers in terms of position correction gain was better compared to verifying according to bone anatomy (p<0.05). Target focused position verifying was the best way for reducing both systematical and random set-up errors.

CONCLUSION
The decrease in standard deviation of systematical and random errors provides statistically meaningful decrease in minimum PTV margins. Keywords : Gold marker; a-Si EPID; elektronic portal imaging; setup errors