Summary
OBJECTIVEThe study aims to evaluate the imaging and positioning accuracy of the BrainLab ExacTrac Dynamic System integrated with the Varian TrueBeam Linear Accelerator for stereotactic radiotherapy.
METHODS
BrainLab ExacTrac Dynamic System, BrainLab Phantom Pointer, and BrainLab Head Phantom were
utilized. Imaging and positioning accuracy were assessed through manual and couch-induced displacement
tests. This article does not contain any studies with human participants or animals performed
by any of the authors. Ethical approval was not required for this study as it was conducted
using phantoms only.
RESULTS
The study evaluated the accuracy of BrainLab ETD and Varian cone-beam computed tomography
(CBCT) systems using a shared reference. Both systems demonstrated sub-millimeter precision, with
translational errors ranging between 0.06 mm (ETD Z-axis) and 0.04 mm (ETD X-axis), and rotational
errors below 0.12°. Maximum observed errors did not exceed 0.3 mm for translational or
0.2° for rotational movements. For combined errors 3D translational accuracy showed no significant
difference p=0.482) between ETD (0.11 mm) and CBCT (0.09 mm). The 6D error magnitude was
slightly lower for ETD (0.14 mm vs. 0.20 mm, p<0.001). Correlation analysis revealed that single-axis
correlations between ETD and CBCT were not significant (p>0.05) except for pitch errors (p=0.023),
which showed moderate correlation. The mean DIFF 6D vector (0.23 mm), representing the positional
offset between ETD and CBCT detections, was significantly larger than both ETD 6D and CBCT
6D errors (p<0.001 and p=0.002, respectively).
CONCLUSION
Both systems offer high precision in detecting positional shifts. The ETD System"s ability to provide continuous,
real-time monitoring enhances patient safety by allowing real-time correction of positioning
errors, making it a valuable tool for stereotactic radiotherapy applications. Further studies with patient
data are needed to validate the clinical application of these findings.
Introduction
Stereotactic radiotherapy (SRT) has significantly advanced the treatment of intracranial and extracranial lesions by enabling the precise delivery of high radiation doses while minimizing exposure to surrounding healthy tissues. The effectiveness of SRT is critically dependent on the accuracy of the imaging and positioning systems used during treatment.[1-3] With the development of advanced imaging technologies such as cone-beam computed tomography (CBCT) and sophisticated linear accelerators, the precision and efficacy of SRT have been substantially enhanced.[4,5] However, the continuous evaluation of these systems' imaging and positioning accuracy is essential to ensure optimal treatment outcomes.In recent years, the integration of imaging systems like the BrainLab ExacTrac Dynamic (ETD) Imaging System with linear accelerators such as the Varian TrueBeam STX has further refined SRT capabilities. These systems provide real-time imaging and positioning adjustments, allowing for highly accurate target localization and dose delivery.[4,6] Despite these technological advancements, rigorous calibration and quality assurance (QA) protocols remain imperative to maintain the precision of these sophisticated systems.
Numerous studies have underscored the importance of regular calibration and QA checks in maintaining the accuracy of SRT systems. The use of phantoms and specialized QA tools, such as the Varian Machine Performance Check (MPC) software and the BrainLab Phantom Pointer, has been instrumental in verifying the alignment and performance of imaging and treatment delivery components.[7,8] Calibrating kV and MV imaging systems, along with ensuring the precise alignment of isocenters, is crucial for accurate radiation dose targeting.[9,10]
This study aims to provide a comprehensive evaluation of the imaging and positioning accuracy of SRT systems using a series of calibration and QA tests. By employing tools such as the BrainLab Phantom Pointer, BrainLab Head Phantom, and various displacement tests, we seek to validate the precision of these systems in clinical settings. Our methodology includes detailed calibration routines for CBCT, MV, and kV imaging systems, as well as performance checks using the Varian MPC software. Additionally, we assess the systems" response to manual and couch-induced displacements to thoroughly examine their accuracy under different conditions.
The findings of this study will contribute to the existing body of knowledge on the importance of stringent QA protocols in SRT. By ensuring that all mechanical and dosimetric settings are accurately calibrated, this research aims to enhance the reliability of stereotactic treatments and improve patient outcomes. This work underscores the critical role of continuous quality assurance in the evolving landscape of radiation oncology, highlighting the necessity of meticulous calibration and evaluation practices to maintain the high standards required for effective cancer treatment.
Methods
Hardware and SoftwareThe study utilized two linear accelerators: the Varian TrueBeam STX Linear Accelerator and the Varian TrueBeam Linear Accelerator (Varian Medical Systems, Palo Alto, CA, USA). For imaging systems, the Brain- Lab ExacTrac Dynamic Imaging System (BrainLab, Munich, Germany) and GE Discovery RT Computed Tomography Simulator (GE Healthcare, Chicago, IL, USA) were employed. Treatment planning was conducted using the Eclipse 16.1 Treatment Planning System (Varian Medical Systems, Palo Alto, CA, USA) and the Elements 2.0 Treatment Planning System (Brain- Lab, Munich, Germany). Quality assurance and calibration were ensured using the BrainLab Phantom Pointer (Ball-Bearing), BrainLab Head Phantom, and ISO Cube Phantom (Standard Imaging, Middleton, WI, USA).
Calibration and Quality Control
Imaging Systems Calibration
Initial calibration routines were performed to enhance
the imaging quality of both linear accelerators. These
included dark field and flood field calibrations, identification
and exclusion of dead detector pixels, and adjustment
of kV collimator jaw settings. Subsequently,
similar calibration processes were carried out for the
CBCT system. The isocenters of the kV and MV imaging
systems were aligned with the isocenter of the
linear accelerator using the Varian IsoCal phantom. By
rotating the gantry through 360 degrees and capturing
images, any deviations of the geometric center from
the device"s central axis were detected and corrected.
Machine Performance Check (MPC)
The MPC involved comprehensive mechanical and dosimetric
tests of the treatment devices. The Varian MPC
software and MPC check phantom were used to conduct
numerous quality control tests, including checks
on gantry, collimator, and couch alignment, as well as
the accuracy of the jaws and MLC positions. These tests
ensured that all calibrations were verified and met established
limits before starting the measurements.
Picket Fence and Winston-Lutz Tests
To verify the positional accuracy of the MLCs and optimize
the success rate of the Hancock Winston-Lutz
test, the picket fence test was conducted. Using the
Eclipse planning system, plans were created at gantry
angles of 0, 90, 180, and 270 degrees, covering all
MLCs. Each plan included a defined irradiation time
of 20 MU for each subfield, ensuring proper alignment
and gaps between MLC pairs.
The Hancock Winston-Lutz test involved creating twelve different areas with 6 MV photon energy, incorporating specific gantry, couch, and collimator angles. Each area was designed to prevent leakage doses by positioning the MLC intersection areas outside the radiation field. The BrainLab phantom pointer (ballbearing) was positioned according to the light field centers at gantry angles of 0 and 90 degrees. Port images obtained from the irradiation were analyzed using the SNC Machine software, and necessary corrections were applied. The device"s isocenter was then radiologically positioned with high precision, and the BrainLab ETD system was calibrated to this point.
<
Evaluation of Imaging and Positioning Accuracy
Before evaluating the imaging and positioning accuracies
of the stereotactic radiotherapy systems, all
mechanical and dosimetric settings of the treatment
devices were verified for accuracy, and necessary adjustments
were made. The isocenters of both systems
were precisely calibrated.
BrainLab Phantom Pointer (Ball-Bearing) Tests
Manual Displacement Test
We moved the ball-bearing (BB) in one direction in 3D
space to a known distance "d" and measured the displacement
(Fig. 1). Keeping the first displacement, we
then moved the BB the same amount in the second direction
and measured again. Subsequently, we moved
the BB to the distance "d" in the third direction and
measured once more. The final coordinate after the
third measurement was (d, d, d). We repeated this process
with negative "d" values, creating a set of six measurements
for each selected distance. We performed this
method using distances of 0.1 mm, 0.5 mm, 1.5 mm,
and 3.0 mm, resulting in a total of 24 measurement sets.
The aim of this method was to utilize varying distances
to test the detection abilities of the systems and to cover
the full 3D space as homogeneously as possible.
Fig. 1. BrainLab phantom pointer (Ball-Bearing).
Couch Displacement Test
Following the manual displacement tests, the BrainLab
Phantom Pointer was repositioned to the isocenter, and displacements were applied using the treatment
couch. These displacements were detected both by the
BrainLab ETD system and the integrated kV and MV
port imaging systems. The responses of the BrainLab
ETD system and the kV and MV port imaging systems
to these displacements were separately analyzed.
Comparison of Positioning Accuracy with BrainLab
ETD and Varian CBCT Using Head Phantom
This part of the study was designed to evaluate the relative
accuracy of the BrainLab ETD and Varian CBCT
systems in detecting displacements applied via a clinically
validated treatment couch. Rather than assuming
CBCT as the gold standard, both systems were independently
assessed using a shared reference. Translational
displacements were validated with a micrometer,
while rotational measurements used the treatment
couch as a consistent reference. This strategy allows a
methodologically unbiased comparison of six degrees
of freedom (6DOF) detection capabilities.
The head phantom was fitted with a BrainLab head and neck positioning mask to ensure stability and reproducibility during measurements (Fig. 2). High-resolution CT images with a slice thickness of 0.625 mm were acquired and imported into the Eclipse treatment planning system. A demo plan with a single field, including kV port and CBCT imaging areas, was created and loaded into both the BrainLab ETD and Varian CBCT systems to enable precise positional comparisons.
Fig. 2. BrainLab head phantom and positioning mask.
Couch Displacement Protocol
Couch displacements were systematically applied
in four predefined distances of 0.1 mm, 0.5 mm,
1.0 mm, and 3.0 mm along each axis (X, Y, Z). For
each distance, displacements were performed in both
positive and negative directions, resulting in eight
measurements per axis. Rotational displacements
(yaw) were also included in the protocol, following
the same sequence of movements and distances. Each
displacement was performed sequentially, first along
the translational axes (X, Y, Z) and then for yaw, ensuring
a comprehensive evaluation of the system"s
6DOF detection capabilities.
To avoid bias present in previous studies, which often compared ETD directly against CBCT while implicitly treating CBCT as the baseline, this study was designed to measure both systems independently against the same source of displacements: the treatment couch. By using the couch as the reference for both ETD and CBCT, the study evaluated their relative performance on an equal basis.
Measurement and Data Recording
For each applied displacement, six values were recorded,
reflecting positional deviations across all degrees
of freedom: translations (X, Y, Z) and rotations (yaw,
pitch, roll). Translational displacements were validated
against a high-precision micrometer, ensuring accuracy
within 0.02 mm. This validation established the
treatment couch as a reliable reference for translational
movements along the X, Y, and Z axes.
To evaluate rotational detection, displacements were applied along the yaw axis, incorporating the inherent reliability of the treatment couch as the reference. Although the rotational accuracy of the couch was not independently validated, its use as a consistent reference allowed for a relative comparison of the BrainLab ETD and Varian CBCT systems. Yaw rotations were included to evaluate the relative performance of the two systems, rather than to establish absolute rotational accuracy.
For each displacement, the 6D error magnitude (E6D) was calculated using the following formula:

where Δi represents the deviation in each degree of freedom. This calculation was performed for each applied displacement to provide a comprehensive measure of positional error for both systems.
Statistical Analysis
All data were analyzed using IBM SPSS Statistics (version
29.0.1, IBM Corp., Armonk, NY, USA). Due to
the small sample sizes and non-normal distribution
of error values (verified via visual inspection and distribution
metrics), nonparametric statistical methods
were used throughout the study. The Wilcoxon
signed-rank test was applied to compare paired error measurements between systems, and correlation
analyses were conducted to evaluate inter-system
agreement across individual and combined axes. Statistical
significance was defined at p<0.05.
All graphical representations were created using the BoxPlotR web-based tool.[11] (http://shiny.chemgrid. org/boxplotr). These boxplots illustrate the distributions and significant differences between the systems and their respective errors, as discussed in subsequent sections.
Results
The following results were obtained from measurements and statistical analyses. Key comparisons are presented for translational and rotational positioning errors across systems.
Micrometer vs ETD Accuracy
ETD system detected all manual displacements of
the Ball-Bearing (BB) phantom (0.10-3.00 mm) with
0.02 mm precision, as measured using the integrated
micrometer at two-digit sensitivity (0.01 mm resolution).
No discrepancies were observed between applied
displacements and ETD-reported values, even at the
smallest tested increment (0.10 mm).
This confirms the system's reliability for subsequent
tests requiring 0.1 mm clinical sensitivity. The results validate
ETD"s capability to resolve submillimeter positional
deviations, ensuring its suitability for high-precision stereotactic
radiotherapy quality assurance protocols.
Couch Displacement Detection
The precision of BrainLab ETD, MV and kV imaging
systems in detecting couch-induced displacements was
evaluated using the Ball-Bearing (BB) phantom. Couch
shifts (0.10-3.00 mm) were applied along all translational
axes (X, Y, Z) using service mode commands (two-digit
precision, 0.01 mm resolution), with displacements
recorded by the ETD and orthogonal MV/kV imaging
systems at clinical one-digit precision (0.1 mm).
The ETD system detected all 30 couch displacements with 0.1 mm maximum error per axis, mirroring its performance in micrometer-based tests. For MV/kV systems, individual axis errors did not exceed 0.1 mm, and 3D vector errors reached a maximum of 0.14 mm (mean: 0.08±0.04 mm). Wilcoxon signed rank tests confirmed that there were no significant differences between MV and kV systems (p>0.05), with errors exhibiting random distribution (no inter-system correlation, r<0.2) (Fig. 3).
These results validate the treatment couch as a reliable reference for submillimeter positional adjustments in clinical workflows. The consistency between the ETD, MV, and kV systems underscores their collective precision for stereotactic radiotherapy applications that require a tolerance of 0.1 mm.
Head Phantom Movements
Translations and Rotations Detected by ETD and
CBCT
The descriptive statistics for positional errors demonstrated
submillimeter precision across both the ETD and CBCT systems. For translational axes, mean errors
ranged between 0.06 mm (ETD Z axis) and 0.04
mm (ETD X axis), with standard deviations <0.07 mm.
Rotational errors were similarly minimal, with mean
deviations <0.12° and standard deviations <0.07°. The
maximum observed errors did not exceed 0.3 mm or
0.2°, underscoring the high precision of both systems.
Although significant differences (p<0.05) were observed for specific axes (e.g., ETD demonstrated smaller errors in Y/Z translations and pitch rotations, CBCT on the X axis and yaw), the absolute error magnitudes were clinically negligible. For example, the mean difference in the Y-axis errors between ETD and CBCT was 0.04 mm, and the pitch rotation discrepancies averaged 0.09°.
For combined errors, 3D translational accuracy showed no significant difference between systems (p=0.482), with mean errors of 0.11 mm (ETD) and 0.09 mm (CBCT). The boxplots for translational errors (X, Y, Z) and rotational vector magnitude (RVM) errors (Fig. 4) further emphasize the sub-millimeter precision of both systems, highlighting minimal variability and similar error ranges across axes.
The 6D error magnitude, incorporating translational and rotational deviations, was slightly lower for ETD (0.14 mm vs. 0.20 mm, p<0.001). The boxplot depicting 6D error magnitude (Fig. 5) demonstrates a statistically significant difference between ETD and CBCT (p<0.001), with ETD achieving lower combined translational and rotational error magnitudes. The correlation analysis showed that 5 of 6 single-axis correlations between ETD and CBCT were not statistically significant (p>0.05), indicating no meaningful agreement. Although, pitch errors were moderately correlated (p=0.023), suggesting some level of agreement for this axis. Despite this single-axis correlation, the overall lack of significant relationships across axes underscores the systems' independent error detection and operational differences.
For the combined 6D error, which integrates translational and rotational deviations, the correlation coefficient was not significant (p=0.653). This highlights the independent detection of 6D errors by the two systems, with no systematic relationship observed across the overall error magnitudes. These findings emphasize the robustness of both systems for stereotactic radiotherapy, with clinically negligible differences in precision. The correlation analysis further supports the conclusion that ETD and CBCT systems independently achieve high accuracy in positional detection, with minor variations likely attributable to system-specific operational characteristics rather than systematic biases.
Comparison of ETD and CBCT Error Magnitudes
Using a Difference Vector
Our study evaluated both the ETD and CBCT systems
against a shared reference, a phantom displaced
by the treatment couch and we showed that both systems
achieve sub-millimeter precision in error detection.
While some prior literature assumes CBCT as the
ground truth, this methodology may lead to inaccurate
estimations of error magnitudes when comparing ETD
detections directly to CBCT. This can exaggerate the
magnitude, as the difference reflects not only the magnitude
but also the directional offset between the two
systems. Conversely, if the vectorial directions of ETD
and CBCT errors align closely, the magnitude may
appear artificially reduced, masking the true extent of
discrepancies. To evaluate the discrepancy, we computed
a difference vector (DIFF-6D), representing the
positional offset between ETD and CBCT detections.
The mean DIFF-6D was 0.23 mm, larger than both
ETD-6D and CBCT-6D. Statistical analysis confirmed
that the DIFF- 6D vector was significantly greater than
the 6D errors of ETD and CBCT systems (p<0.001 and
p=0.002 respectively) (Fig. 6).
These results highlight the potential overestimation of ETD errors when compared directly to CBCT without an independent reference. While CBCT remains a clinically indispensable tool due to its anatomical visualization capabilities, our findings emphasize the importance of unbiased methodologies for evaluating detection capabilities of imaging systems.
Discussion
Due to the use of high doses with a small number of fractions and small PTV margins in SRS/SBRT treatment techniques, correct patient positioning is extremely important to prevent complications in surrounding tissues and organs at risk and to ensure tumor control. [1,2,4,9,10] Correct patient positioning (setup) before treatment is essential to reduce doses to organs at risk while delivering the prescribed dose to the tumor.[12,13] A 1-2 mm margin (usually 1 mm) is used for GTV-PTV in brain tumors. It has been reported that larger margins may cause complications such as radiation-induced brain necrosis. V12Gy should be kept below 5 to 10 cc to reduce brain necrosis.[3,14,15] In SRS/SBRT treatment techniques, precise patient positioning and/or sub-millimeter accuracy monitoring is required due to small PTV margins and potential patient movements.[1]Since thermoplastic masks do not completely eliminate potential movements during stereotactic treatments, patient motion of varying magnitudes may occur between each arc of dose delivery, which can lead to overdosing of normal tissues near the target.[5,16] Although CBCT, widely used before treatment in IGRT, provides three-dimensional anatomical images, it cannot track intrafractional/interfractional movement during treatment. A real-time intrafractional patient tracking system capable of optical surface scanning has been developed, but these systems lack internal anatomy information. In our clinic, the "ExacTrac Dynamic (ETD)" system is used, which includes X-ray tubes and flat panel detectors. This system monitors changes in patient anatomy or positioning independently of the treatment device's parameters and corrects any discrepancies. The ETD system also offers 3D optical surface scanning and thermal monitoring, providing verification methods beyond treatment devices alone.[4]
Setup and intrafraction movements can be corrected using the ExacTrac X-ray 6D imaging system, allowing for reduced GTV-PTV margins.[3] The ExacTrac system detects uncertainties caused by translational and rotational movements, allowing for verification and correction before and during treatment. It offers advantages over CBCT, such as shorter image acquisition times and lower patient radiation exposure.[5] For example, one study found a simultaneous radiation dose of 2.0 mSv with ExacTrac compared to 14.0 mSv with CBCT.[17]
CBCT is essential in all radiotherapy treatments due to its superior visualization of anatomical structures and soft tissues, offering 3D and transaxial images before treatment. Geometric uncertainties are detected more clearly in SRS/SBRT through integrating multiple imaging techniques.[18] Proper and accurate IGRT techniques must be used to detect intrafraction and interfraction errors.[13] Additionally, correct positioning of the radiation isocenter during gantry, collimator, and couch rotation depends on the use of accurate IGRT techniques.[1]
In SRS for multiple brain metastases, correct isocenter positioning during treatment is crucial to ensure precise dose delivery to all targets.[3] Rotational uncertainties in single-isocenter SRS techniques can significantly impact target coverage for lesions located far from the isocenter.[1]
To determine the accuracy of the ExacTrac system, the first step is to verify patient positioning accuracy relative to the linear accelerator isocenter. Secondly, imaging and couch corrections must be confirmed. [5] Calibration of the ExacTrac isocenter is regularly performed using a special 10x10 cm² phantom with a reflective sphere, ensuring alignment with the linear accelerator isocenter.
A Winston-Lutz (WL) phantom, with a small metal sphere inside, is used to determine the ExacTrac X-ray isocenter relative to the radiation isocenter. The difference between the sphere"s center projection and the treatment area center should be <0.7 mm.[10] Arp & Carl[8] found deviations between the outer laser isocenter and the ExacTrac isocenter ranging from 0.21 to 0.42 mm, while the difference between the laser isocenter and radiation isocenter ranged from 0.37 to 0.83 mm. The three-dimensional deviation between the radiation isocenter and the ExacTrac center was between 0.31 and 1.07 mm. Verellen et al.[19] reported a deviation of 0.24 mm at the radiation isocenter using hidden target tests.
In our study, we determined the radiation isocenter using similar methods. After examining the 2D images taken at different gantry, collimator, and couch angles, the radiation isocenter was determined by shifting the metal sphere on the ETD phantom with micrometer precision. The BB phantom provided with the ETD system was used to ensure sub-millimeter accuracy in detecting the isocenter, showing minimal deviation. The ETD system performed accurately in all measurements taken at 90 different points, detecting the BB without error, with maximum deviations of 0.1 mm from the radiation isocenter.
Verbakel et al.[7] reported a deviation of 0.3 mm in each direction after randomly positioning an Alderson head phantom 31 times and performing automatic 6D couch shifts using ExacTrac. Intrafraction movement was measured at 0.35±0.21 mm (maximum 1.15 mm) for 46 patients. Wurm et al.[20] found an average translational and rotational setup error of 0.31 mm and 0.26°, respectively. Jani et al.[21] reported a positional accuracy of <1 mm with ExacTrac in 94 patients, emphasizing the importance of real patient data in understanding intrafractional motion. Gevaert et al.[9] validated the ExacTrac/ Novalis Body system on an anthropomorphic phantom, reporting rotational setup errors with longitudinal precision of 0.09°±0.06° and lateral precision of 0.02°±0.07°.
When comparing ExacTrac and CBCT in terms of setup errors, ExacTrac showed more accurate results. Chow et al.[4] compared the positional shifts detected by ExacTrac"s optical surface thermal monitoring (EXTD-thml) with stereoscopic X-ray (EXTD-X- ray) and CBCT. The translational differences between Exac- Trac-thml and CBCT were 0.57±0.41 mm for the cranial phantom and 0.66±0.40 mm for the pelvic phantom, while the rotational differences were less than 0.7°.
Kojima et al.[22] reported translational errors of less than 1 mm and rotational errors of less than 1°. Ma et al.[23] determined setup errors between EXTD-Xray and CBCT using a cranial phantom, finding translational differences of <0.5 mm and rotational differences of <0.2°. In another study, the setup differences between the BrainLab ExacTrac X-ray system and CBCT for intracranial stereotactic radiotherapy were reported as <1.01 mm for online matching and <0.82° for rotational differences.[13] Ma et al.[23] found translational differences of <0.5 mm for phantoms and <1.5 mm for patients, and rotational differences of <0.2° for phantoms and <1.0° for patients. Chang et al.[2] evaluated setup errors between ExacTrac X-ray and CBCT in spinal non-invasive stereotactic body therapy (SBRT), finding translational differences of <1.0 mm and rotational differences of <1.0° for phantoms, while patient data showed differences of <2.0 mm and <1.5°.
Spadea et al.[24] reported intrafraction variability between ExacTrac X-ray and CBCT using IR optical localization, with localization differences of 0.3±0.3 mm and X-ray image differences of 0.9±0.8 mm. Takakura et al.[25] found a positional correction accuracy of 0.07±0.22 mm for a head and neck phantom, while the overall geometric accuracy of the Novalis ExacTrac system was reported as 0.31±0.77 mm.
Li et al.[6] found setup error differences between CBCT and BrainLab ExacTrac X-ray were less than 0.4 mm in the translational direction and 0.5° in the rotational direction. Tanyi et al.[26] reported deviations between ExacTrac and 4D CBCT of 0.40±0.72 mm in the vertical direction and -0.06±0.67 mm in the lateral direction from 245 patient matchings, with rotational differences within 0.7°. Montgomery and Collins[5] reported submillimeter reproducibility for ExacTrac based on hidden target tests.
Since fixed phantoms were used in our study, future studies will compare the performance of these systems with real patient data to further evaluate their clinical applicability.
Conclusion
Unlike previous studies that rely on CBCT as the reference standard, this study evaluated the accuracy of BrainLab ExacTrac Dynamic and Varian CBCT systems using an independent reference, a phantom displaced by a micrometer and the treatment couch, to eliminate bias. Both systems demonstrated sub-millimeter precision for translational errors and sub-degree accuracy for rotational deviations, confirming their suitability for high-precision stereotactic radiotherapy. We directly compared 6D error detection accuracy of ETD and CBCT systems against an external reference. Significant differences in 6D error magnitudes were observed between ETD and CBCT, with ETD achieving slightly lower combined errors. Correlation analysis revealed no meaningful agreement for single-axis errors, underscoring the independent error detection capabilities of the systems. The DIFF6D vector, representing positional offsets between the systems, highlighted potential discrepancies when CBCT is assumed as the ground truth, emphasizing the importance of unbiased evaluation methodologies.While CBCT remains indispensable for anatomical visualization in clinical workflows, this study demonstrates that both ETD and CBCT independently achieve the precision required for stereotactic radiotherapy applications. Future research should further investigate system performance in dynamic clinical scenarios especially in pediatric patients to validate these findings.
Conflict of Interest Statement: The authors declare no direct conflict of interests.
Funding: This study was supported by the Ege University Scientific Research Projects Coordination Unit under Project No. 21799. The authors would like to express their gratitude to Ege University for their financial and institutional support.
Use of AI for Writing Assistance: No AI technologies utilized.
Author Contributions: Concept - D.Y., E.T., Y.Z.H., N.O., S.K.; Design - D.Y., E.T., Y.Z.H., N.O., S.K.; Supervision - D.Y., E.T., Y.Z.H., N.O., S.K.; Data collection and/ or processing - E.T., Y.Z.H.; Data analysis and/or interpretation - D.Y., E.T., Y.Z.H.; Literature search - D.Y., E.T.; Writing - D.Y., E.T.; Critical review - D.Y., E.T., Y.Z.H., N.O., S.K.
Peer-review: Externally peer-reviewed.
References
1) Eder MM, Reiner M, Heinz C, Garny S, Freislederer P,
Landry G, et al. Single-isocenter stereotactic radiosurgery
for multiple brain metastases: Impact of patient
misalignments on target coverage in non-coplanar
treatments. Z Med Phys 2022;32(3):296-311
2) Chang Z, Wang Z, Ma J, O"Daniel JC, Kirkpatrick J, Yin
FF. 6D image guidance for spinal non-invasive stereotactic
body radiation therapy: Comparison between
ExacTrac X-ray 6D with kilo-voltage cone-beam CT.
Radiother Oncol 2010;95(1):116-21
3) Minniti G, Capone L, Alongi F, Figlia V, Nardiello B,
El Gawhary R, et al. Initial experience with single-isocenter
radiosurgery to target multiple brain metastases
using an automated treatment planning software:
Clinical outcomes and optimal target volume margins
strategy. Adv Radiat Oncol 2020;5(5):856-64
4) Chow VUY, Cheung MLM, Kan MWK, Chan ATC.
Shift detection discrepancy between ExacTrac Dynamic
system and cone-beam computed tomography.
J Appl Clin Med Phys 2022;23(5):e13567
5) Montgomery C, Collins M. An evaluation of the
BrainLAB 6D ExacTrac/Novalis Tx system for image-
guided intracranial radiotherapy. J Radiother
Pract 2017;16:326-33
6) Li J, Shi W, Andrews D, Werner-Wasik M, Lu B, Yu Y,
et al. Comparison of online 6 degree-of-freedom image
registration of Varian TrueBeam cone-beam CT
and Brainlab ExacTrac x-ray for intracranial radiosurgery.
Technol Cancer Res Treat 2017;16(3):339-43
7) Verbakel WFAR, Lagerwaard FJ, Verduin AJE, Heukelom
S, Slotman BJ, Cuijpers JP. The accuracy of frameless
stereotactic intracranial radiosurgery. Radiother
Oncol 2010;97(3):390-4
8) Arp DT, Carl J. EXACTRAC x-ray and beam isocentres-
what"s the difference? Med Phys 2012;39(3):1418-
23
9) Gevaert T, Verellen D, Tournel K, Linthout N, Bral S,
Engels B, et al. Setup accuracy of the Novalis ExacTrac
6DOF system for frameless radiosurgery. Int J Radiat
Oncol Biol Phys 2012;82(5):1627-35
10) Almeida TVR, Junior ALC, Piedade PA, Da Silva CM,
Marins P, Almeida CM, et al. Analysis of translational
errors in frame-based and frameless cranial radiosurgery
using an anthropomorphic phantom. Radiol Bras
2016;49(2):98-103
11) Spitzer M, Wildenhain J, Rappsilber J, Tyers M. Box-
PlotR: A web tool for generation of box plots. Nat
Methods 2014;11(2):121-2
12) Lewis BC, Snyder WJ, Kim S, Kim T. Monitoring
frequency of intra-fraction patient motion using the
ExacTrac system for LINAC-based SRS treatments. J
Appl Clin Med Phys 2018;19(3):58-63
13) Oh SA, Park JW, Yea JW, Kim SK. Evaluations of
the setup discrepancy between BrainLAB 6D Exac-
Trac and cone-beam computed tomography used
with the imaging guidance system Novalis-Tx for
intracranial stereotactic radiosurgery. PLoS ONE
2017;12(5):e0177798
14) Korytko T, Radivoyevitch T, Colussi V, Wessels BW,
Pillai K, Maciunas RJ, et al. 12 Gy gamma knife radiosurgical
volume is a predictor for radiation necrosis in
non-AVM intracranial tumors. Int J Radiat Oncol Biol
Phys 2006;64(2):419-24
15) Minniti G, Clarke E, Lanzetta G, Osti MF, Trasimeni
G, Bozzao A, et al. Stereotactic radiosurgery for brain
metastases: Analysis of outcome and risk of brain radionecrosis.
Radiat Oncol 2011;6:48
16) Hoogeman MS, Nuyttens JJ, Levendag PC, Heijmen
BJM. Time dependence of intrafraction patient motion
assessed by repeat stereoscopic imaging. Int J Radiat
Oncol Biol Phys 2008;70(2):609-18
17) Jin JY, Ryu S, Faber K, Mikkelsen T, Chen Q, Li S, et
al. 2D/3D image fusion for accurate target localization
and evaluation of a mask based stereotactic system in
fractionated stereotactic radiotherapy of cranial lesions.
Med Phys 2006;33(12):4557-66
18) Zollner B, Heinz C, Pitzler S, Manapov F, Kantz S,
Rottler MC, et al. Stereoscopic x-ray imaging, cone
beam CT, and couch positioning in stereotactic radiotherapy
of intracranial tumors: Preliminary results
from a cross-modality pilot installation. Radiat Oncol
2016;11(1):158
19) Verellen D, Soete G, Linthout N, Van Acker S, De
Roover P, Vinh-Hung V, et al. Quality assurance of
a system for improved target localization and patient
set-up that combines real-time infrared tracking
and stereoscopic x-ray imaging. Radiother Oncol
2003;67(1):129-41
20) Wurm RE, Erbel S, Schwenkert I, Gum F, Agaoglu D,
Schild R, et al. Novalis frameless image-guided noninvasive
radiosurgery: Initial experience. Neurosurgery
2008;62(5):A11-8
21) Jani SK, Weinstein G, Shah N, Volpp P, Glazebrook
S, Pothilat J, et al. Performance of an ExacTrac image
guidance system in mask-based cranial radiotherapy
on a classic Novalis machine: A five-year experience.
Int J Radiat Oncol Biol Phys 2010;78(3):S765-6
22) Kojima H, Takemura A, Kurokawa S, Ueda S, Noto
K, Yokoyama H, et al. Evaluation of technical performance
of optical surface imaging system using conventional
and novel stereotactic radiosurgery algorithms.
J Appl Clin Med Phys 2021;22(2):58-68
23) Ma J, Chang Z, Wang Z, Wu QJ, Kirkpatrick JP, Yin FF.
ExacTrac x-ray 6 degree-of-freedom image-guidance
for intracranial non-invasive stereotactic radiotherapy:
Comparison with kilo-voltage cone-beam CT. Radiother
Oncol 2009;93(3):602-8
24) Spadea MF, Tagaste B, Riboldi M, Preve E, Alterio D,
Piperno G, et al. Intra-fraction setup variability: IR optical
localization vs x-ray imaging in a hypofractionated
patient population. Radiat Oncol 2011;6:38




