Summary
Surface-guided radiation therapy (SGRT) has gained wide popularity across radiation oncology community due to its non-radiographic characteristic and real-time motion monitoring capability. Nevertheless, it has not yet gained its full potential in routine clinical practice. Implementing SGRT system into the clinical practice requires not only the definition of steps in clinical workflow, but also establishment of the comprehensive quality assurance (QA) program including commissioning, acceptance and periodic QA test to facilitate a safe, and efficient use of SGRT system in clinical settings. This review focuses on the latest recommendation of American Association of Physicists in Medicine and European Society for Radiotherapy and Oncology guidelines about the implementation of comprehensive QA program for SGRT system.Introduction
Surface-guided radiation therapy (SGRT) has emerged as a special form of image-guided radiation therapy (IGRT). Since its first introduction as a useful IGRT tool, many researcher have explored the feasibility of SGRT system for patient positioning, realtime motion management, four-dimensional imaging for motion tracking and threshold gating.[1-8] In several studies, it was proved to improve initial patient positioning by correcting posture differences before online imaging.[9,10] Nevertheless, after initial patient setup with SGRT, online imaging modalities such as planar imaging (kV or MV) and cone-beam computed tomography still needs to be performed in many anatomical sites especially located in thorax and abdomen since the changes in internal motion remain undetected through surface scanning with current technology.[9] However, up to that time, limited guidelines and the complexity of the clinical settings have led to diverse patterns of practice between the clinics. In 2019, Padilla et al.[11] conducted an electronic survey under the auspices of American Association of Physicists in Medicine (AAPM) Task Group Report 302 (TG-302) to identify the necessity of formal guidance and to gain more insight on prevalence of the SGRT system in USA, length of its use, existing recommendation for commissioning procedures and clinical implementation. According to questionnaire, 36% of the users (n=86) only followed the vendor's guidelines and 49.1% of the respondents (n=115) used more than one reference during commissioning. In terms of the question about the use of any end-toend (E2E) test verification approaches, 12% of the users (n=28) response this question as "No" and 14.1% of the respondents (n=33) do not know whether they performed any E2E test, or not. Similarly, in 2022, another international survey was conducted with the collaboration of European Society for Radiotherapy and Oncology (ESTRO) and AAPM to provide an overview about the current status of SGRT in clinical practice with a focus on the user's experience in terms of implementation, commissioning, periodical quality assurance (QA), training, and clinical workflow.[12] According to results of the survey, clinical implementation of the SGRT systems was predominantly based on the vendor's recommendation. Indeed, 94% of the respondents (n=132) primarily followed vendor's guidelines during clinical implementation, commissioning and periodical QA. About 42% of the participant (n=59) used two different sources and only 19% (n=27) used at least three different sources including vendor's guidelines and published studies in the literature or peer-to-peer consultation. In addition, 54% of the respondents (n=76) exclusively used QA tools provided by the vendors during the commissioning and periodical QA tests. About 44% of the users (n=62) preferred to use vendor-provided phantom in combination with the third party commercially available phantoms (n=34) and/or with in-house phantoms (n=28). However, 8% of the respondents (n=12) reported the use of only either third-party commercial phantoms (n=6) or adapted in-house phantoms (n=6) instead of vendor-provided phantoms. According to results of both surveys, it was strongly emphasized that consensus guidelines on SGRT are needed for standardization in clinical practice since the use of different techniques during implementation, commissioning and periodical QA test may cause a systematic errors in patient setup and monitoring. Recently, two different guidelines were published by AAPM, called as TG-302[1] building on the TG-147[13] report, and ESTRO-ACROP[2] to expedite its safe adaptation in clinical practice. AAPM TG-302 also referred other guidelines such as AAPM TG-76[14] and AAPM TG- 142[15] for several QA tests. Although both of these guidelines (AAPM TG-302 and ESTRO-ACROP) were comprehensive and informative, there are still several differences in terms of suggested parameters and tolerance values that need to be considered during QA program including acceptance, commissioning, and periodical QA tests. We aimed to compare both guidelines in terms of recommended parameters based on system specification (simulation room vs. treatment rom, C-arm vs. ring gantry, photon vs. particle etc.), type of tests and tolerances/specifications during the acceptance, commissioning, and periodical QA tests. In addition, phantom selection criteria for SGRT QA and current challenges in SGRT QA were discussed in detail.
QA PROGRAM FOR SGRT
Acceptance Test
The acceptance process need to include all required tests
including static/dynamic localization accuracy, spatial
reproducibility and drift to illustrate the safe operation
and proper functionality of the SGRT system with the
integrated treatment or simulation platform. In most
cases, the acceptance test document is provided by the
vendor and it may not include all necessary tests that
need to be checked. However, it is important to keep in
mind that the acceptance procedure is an integral part
of the purchasing process to ensure whether the product
or solution meet the clinical need, or not. Therefore,
primary responsible person, generally qualified medical
physicist expert, needs to be familiar with the fundamental
or basic tests recommended in the commissioning
and if these tests are not included in the vendor's
acceptance documents, it is generally recommended to
negotiate with the vendor to perform these tests during
acceptance. According to AAPM TG-302, vendor's
recommendation and other AAPM reports such as TG-
142, TG-147, and TG-76 need to be followed together
for checking the localization accuracy and reproducibility
of the system. In addition, safe operation and proper
functionality of the system with all other unit interface,
including imaging system (if necessary), treatment machine,
treatment planning system, data transfer and information
system, and need to be validated as described
in Table 1. In contrast to TG-302, ESTRO-ACROP
guideline provides a more detailed information about
the description of each parameters.[2] Moreover, all
tests are categorized with respect to importance level
(x- mandatory, o-optional, pass-within Vendor's system
specifications), type of systems (computer tomography
[CT], closed-bore linac, C-arm linac and particle therapy),
and subgroups for each suggested parameters.
Table 1 Recommended parameters for acceptance tests in ESTRO-ACROP and AAPM guidelines
Commissioning
The commissioning of the SGRT system is a substantial
part of the comprehensive QA program before implementing
it into clinical practice. This part also includes
measuring the system accuracy/precision and determining
system limitations for all clinically relevant
scenarios. Since the commissioning data are accepted
as a reference for future measurement, all tests need to
be reproducible to assess the consistency of the system
performance over the period of time for periodical QA
tests or for later measurements after maintenance and
repair of the system. In addition, according to AAPM
TG-147 recommendation, commissioning test need to be repeated in case of special situation, ranging from
major upgrade and power outages to earthquake and
building vibration, to check the stability of the system
before using it in clinical practice. All suggested parameters
for system commissioning in AAPM TG-302
and ESTRO-ACROP guidelines are summarized in
Table 2. Some specification and tolerance values were
tightened in ESTRO-ACROP guidelines and new tests
were described based on the availability of new technologies
and updated clinical needs.
Periodic QA Program
The main goal of the periodical QA program is to ensure
about the stability of the system over a period of
time (e.g., daily, weekly, monthly, and annually) and
to catch the unexpected errors or changes in system
performance due to the many factors such as component
failure, machine malfunction or aging of the
system component. ESRTO-ACROP guideline also
recommended to start with a higher frequency and
higher number of tests until the RT team feel more
comfortable about the stability of the system based on
the test outcome preferably including a failure modes
and effective analysis specific to the clinic. In addition,
ESTRO-ACROP guideline reported the list of failure
modes and potential errors in SGRT workflows with
possible solutions. Similar to acceptance and commissioning
part, ESTRO-ACROP guideline provides more
comprehensive periodic QA program compared to
AAPM TG-302 and TG-147 recommendations as presented
in Table 3. Detailed information and description
of each test are also provided in both AAPM TG-
147 and supplement of ESTRO-ACROP guidelines.
Table 3 Periodical QA tests for SGRT
QA Phantoms for SGRT
SGRT requires dedicated QA phantoms with specific
properties (e.g., color, reflectivity, texture, and topography)
that make it accurately trackable. Although some
commercially available SGRT systems allow the user
to change imaging parameters (e.g., camera light and
exposure time) for capturing surface information from
the bodies/phantoms with variety skin/surface tones,
opaque/matte and light colored phantoms yields the
best monitoring results during QA due to the better reflection
characteristic for the projected light pattern. In
fact, the use of SGRT system in variety skin tones, especially
in case of dark skin, is still one of the challenging
issues to consider in clinical practice. However, ESTROACROP
guidelines recommended to check localization
accuracy of the SGRT system with both light- and darktoned
phantoms when it is possible, especially in clinics
where a larger proportion of patients with darker skin tones are treated. In addition, it needs to be taken in to
account that if the surface of the phantom is shiny, it
may also cause numerous or unwanted reflection pattern
of the projected light. Therefore, in case of necessity,
it is generally recommended to cover the phantom
surface with a paint coat or light colored tape. In addition
to color and reflectivity properties, topography
and texture of the QA phantom may significantly affect
the result of the QA tests. Indeed, in case of insufficient
topography, it is difficult to discern position or motion
of the phantom during the check of localization accuracy
of SGRT system. To overcome this issue, vendors
provide dedicated phantoms that mimic anatomical
surfaces such as the head, leg, or breast. In many clinics,
homemade Styrofoam phantom with a different topography
is also used as an inexpensive way of 3D surface
phantom for SGRT. However, we need to be careful
that Styrofoam with expanded polystyrene beads may
cause uncertainties due to the abundance of texture and
the projected light pattern may not be identified correctly.
Therefore, smoot foam phantoms satisfying the
outlined recommendation in both ESTRO-ACROP and
AAPM guidelines can be also good alternative to commercially
available phantoms. Several types of commercially
available phantoms were also demonstrated in
AAPM TG-302 and ESTRO-ACROP guidelines.
Challenges in SGRT QA
As also defined in AAPM TG-302, there are still several
major issues that cause in uncertainties during both
QA and clinical practice of SGRT. For instance, the use
of DICOM based surface structure generated from CT
imaging is considered as the one of the challenging issue
for accurate localization of the phantom/body. In fact,
many parameters (e.g., CT voxel size, scan speed, respiratory
phase effect for moving phantom/surface, Hounsfield
unit threshold for surface segmentation, and image
quality) can significantly affect the topography of
reference body surface ant it may cause a systematic bias
during localization. Similarly, the size and the shape of
the selected region-of-interest for surface tracking can
also affect the response of the system during QA. In
addition to these parameters, the tracking accuracy of
the SGRT system can decreases when the component of
treatment unit (e.g. gantry head and kV imaging arms)
occlude the SGRT cameras, especially in non-coplanar
treatment techniques with couch angle. Therefore, all
these parameters need to be checked for different scenarios
to evaluate the impact of defined issues on the
tracking and localization accuracy of the implemented
SGRT system before using in clinical practice.
Conclusion
AAPM TG-302 mainly focused on the implementation of SGRT in C-arm linac. However, the use of SGRT system is also getting widespread in other platforms (like closed-bore linac, robotic gantry system, particle therapy, and CT simulator). Therefore, as also emphasized ESTRO-ACROP guidelines, each system need to have a different parameters and corresponding tolerance values for acceptance, commissioning, and routine QA. In terms of this aspect, an ESTRO-ACROP guideline is more comprehensive than AAPM TG-302. Nevertheless, AAPM TG-302 provides more detailed information about the phantom selection criteria and QA issue unique to SGRT and possible solution for these issues. Therefore, both of these reports need to be used together during the implementation of QA program in clinical settings.Peer-review: Externally peer-reviewed.
Conflict of Interest: I have no conflict of interest.
Financial Support: None declared.
References
1) Al-Hallaq HA, Cerviño L, Gutierrez AN, Havnen-
Smith A, Higgins SA, Kügele M, et al. AAPM task
group report 302: Surface-guided radiotherapy. Med
Phys 2022;49(4):e82-e112.
2) Freislederer P, Batista V, Öllers M, Buschmann M,
Steiner E, Kügele M, et al. ESTRO-ACROP guideline
on surface guided radiation therapy. Radiother Oncol
2022;173:188-96.
3) Kügele M, Mannerberg A, Nørring Bekke S, Alkner S,
Berg L, Mahmood F, et al. Surface guided radiotherapy
(SGRT) improves breast cancer patient setup accuracy.
J Appl Clin Med Phys 2019;20(9):61-8.
4) González-Sanchis A, Brualla-González L, Fuster-
Diana C, Gordo-Partearroyo JC, Piñeiro-Vidal T, García-
Hernandez T, et al. Surface-guided radiation therapy
for breast cancer: more precise positioning. Clin
Transl Oncol 2021;23(10):2120-6.
5) Mannerberg A, Kügele M, Hamid S, Edvardsson A,
Petersson K, Gunnlaugsson A, et al. Faster and more accurate
patient positioning with surface guided radiotherapy
for ultra-hypofractionated prostate cancer patients.
Tech Innov Patient Support Radiat Oncol 2021;19:41-5.
6) Li Z, Xiao Q, Li G, Wu X, Zhang Y, Wang G, et al.
Performance assessment of surface-guided radiation
therapy and patient setup in head-and-neck and breast
cancer patients based on statistical process control.
Phys Med 2021;89:243-9.
7) Hoisak JDP, Paxton AB, Waghorn BJ, Pawlicki T. Surface
guided radiation therapy. Boca Raton: CRC Press; 2020.
8) Freislederer P, Kügele M, Öllers M, Swinnen A, Sauer
TO, Bert C, et al. Recent advances in surface guided
radiation therapy. Radiat Oncol 2020;15(1):187.
9) Batista V, Meyer J, Kügele M, Al-Hallaq H. Clinical
paradigms and challenges in surface guided radiation
therapy: Where do we go from here? Radiother Oncol
2020;153:34-42.
10) Bert C, Metheany KG, Doppke KP, Taghian AG, Powell
SN, Chen GT. Clinical experience with a 3D surface
patient setup system for alignment of partial-breast
irradiation patients. Int J Radiat Oncol Biol Phys
2006;64(4):1265-74.
11) Padilla L, Havnen-Smith A, Cerviño L, Al-Hallaq HA. A
survey of surface imaging use in radiation oncology in the
United States. J Appl Clin Med Phys 2019;20(12):70-7.
12) Batista V, Gober M, Moura F, Webster A, Oellers M,
Ramtohul M, et al. Surface guided radiation therapy:
An international survey on current clinical practice.
Tech Innov Patient Support Radiat Oncol 2022;22:1-8.
13) Willoughby T, Lehmann J, Bencomo JA, Jani SK, Santanam
L, Sethi A, et al. Quality assurance for nonradiographic
radiotherapy localization and positioning
systems: report of Task Group 147. Med Phys
2012;39(4):1728-47.