TURKISH JOURNAL OF ONCOLOGY 2024 , Vol 39 , Num 1
Investigating the Influence of Minimum Segment Width in Volumetric-modulated Arc Planning of Prostate Cancer
Nidhi JAIN1,Alok KUMAR2,Ashok KUMAR1
1Department of Physics, Amity University, Noida-India
2Department of Radiation Oncology, Netaji Subhas Chandra Bose Cancer Hospital, Kolkata-India
DOI : 10.5505/tjo.2023.4137

Summary

OBJECTIVE
This study aims to investigate the impact of the minimum segment width on the planning outcomes of volumetric-modulated arc therapy (VMAT) in patients with prostate cancer and find the optimum value(s) for this parameter.

METHODS
A retrospective analysis was conducted on 12 patients with prostate cancer who underwent VMAT treatment. For every patient, four treatment plans were created using different values of MSW (0.5 cm, 1.0 cm, 1.5 cm, and 2.0 cm). Other optimization parameters and objective constraints were kept the same across every case. Several dosimetric parameters were evaluated, including target coverage (Dmean - Mean dose to the planning target volumes [PTV], Dmax - Maximum dose to the PTV, conformity index, homogeneity index) and dose to the organ at risk. In addition, delivery efficiency metrics such as the number of control points, monitor units, and treatment time were assessed. Statistical analyses were performed using Wilcoxon signed-rank test.

RESULTS
Narrower segments (MSW0.5) yielded improved PTV coverage and conformity, while wider segments (MSW2.0) led to faster treatment delivery but compromised dosimetric parameters. There was no statistically significant difference between MSW0.5 and MSW1.0 (p>0.05) while the other MSW values showed statistically significant differences (p<0.05).

CONCLUSION
Based on the analysis of the plan quality and delivery efficiency, an MSW value of 1.0 cm exhibits optimal features in prostate cancer treatment plans. Further investigation with a larger number of patients and assessment of clinical outcomes is necessary to validate this conclusion.

Introduction

Prostate cancer is one of the most prevalent cancers among men, with a significant impact on their health and quality of life.[1] The treatment of prostate cancer often involves radiation therapy, which plays a crucial role in eradicating cancer cells and reducing the risk of recurrence.[2] In recent years, advancements in radiation therapy techniques have greatly improved treatment outcomes, allowing for more precise targeting of tumor volumes while minimizing radiation exposure to healthy surrounding tissues.[3]

One such technique that has gained considerable attention is volumetric-modulated arc therapy (VMAT). VMAT is an advanced form of intensity-modulated radiation therapy that delivers radiation through a continuous arc of motion around the patient. This technique utilizes dynamic modulation of multileaf collimators and gantry rotation to shape the radiation beam precisely, conforming to the tumor"s three-dimensional shape. VMAT offers several advantages over conventional radiation therapy techniques, including shorter treatment times, improved dose conformity, and reduced radiation exposure to healthy tissues.[4,5]

The planning process in VMAT involves the optimization of various parameters to achieve the desired treatment goals. One critical parameter in VMAT planning is the minimum segment width (MSW), which determines the width of the individual radiation beam segments used during treatment delivery. The MSW directly influences the treatment plan"s quality and efficiency by impacting factors such as dose distribution, target coverage, organ sparing, and treatment delivery time.[6] VMAT plan with a higher value of MSW has fewer monitoring units (MU), less delivery time (PDT), and more delivery efficiency than VMAT plan with a lower MSW.[6-11]

Some studies have demonstrated the superiority of the VMAT plan with lower MSW over the plan with higher MSW.[12,13] Several studies have suggested using an MSW of 1.0 cm for the VMAT plan as compared to an MSW of 0.5 cm.[7-9] Therefore, this study aims to investigate the influence of the MSW on the planning outcomes of VMAT in prostate cancer patients. By systematically varying the MSW value, we will evaluate its impact on various dosimetric parameters, including target coverage, dose conformity, organ atrisk sparing, and treatment delivery efficiency metrics. The findings from this study will contribute to a better understanding of the role of MSW in VMAT planning for prostate cancer and help identify the optimal MSW value(s) that can maximize treatment efficacy while minimizing treatment time and potential side effects.

Methods

Patient Selection
The study included twelve patients (aged between 55 and 68 years) who were diagnosed with prostate cancer and received VMAT treatment at our hospital from January 2022 to November 2022. To ensure accurate treatment, all patients were positioned in the supine position and immobilized using a 4-clamp thermoplastic pelvis mask. A Siemens computed tomography (CT) simulator was used to perform scans, with a slice thickness of 3 mm.

The CT images obtained from the scans were then reconstructed and imported into the Monaco planning system version 5.51.10. The planning target volume (PTVP) encompassed the prostate tumor and seminal vesicles, with a 5-mm margin on all sides except for a 3 mm margin posteriorly. The delineation of PTVLN, which included the pelvic lymph nodes, was performed by experienced oncologists following the institute protocol.

In addition, several organs at risk (OAR) structures were delineated, including the bladder, rectum, femoral heads, cauda equina, sigmoid colon, and bowel bags. These structures were outlined to ensure their protection during the treatment planning process.

Treatment planning
VMAT plans were designed for all patients using the Monte Carlo (MC) algorithm in the Monaco Treatment Planning System (TPS). The Elekta Synergy Linear Accelerator with a 6MV X-ray photon beam was utilized to deliver the plans. A dual arc of 360° rotation was employed for each case, clockwise from 181° to 179°. During gantry rotation, the collimator angle was set to 0° based on the patient"s anatomy. The MC algorithm had a statistical uncertainty of 3% per control point, and the final dose calculation used a 3 mm resolution for the calculation grid. Each plan consisted of a maximum of 180 control points (CP).

Four VMAT plans were generated, namely MSW0.5, MSW1.0, MSW1.5, and MSW2.0. These plans had corresponding MSWs of 0.5 cm, 1.0 cm, 1.5 cm, and 2.0 cm, respectively. The remaining parameters and cost functions were kept constant. The prescription dose for PTVP was 68Gy delivered in 25 fractions, while for PTVLN, it was 45Gy in 25 fractions. Table 1 displays the cost functions.

Table 1: The optimization cost functions of VMAT plans for prostate cancer

Plan Evaluation
The dosimetric indices used to compare the different MSW prostate plans included the homogeneity index (HI), conformity index (CI), maximum dose of the target volume, MUs, CP, and the dose volume histogram (DVH) parameters concerning OARs. The CI and HI were calculated as follows:

CI= (TVPI)2/(TVxVPI)

HI = (D5%)/D95%

In the above equations, TVPI represents the target volume receiving the prescription dose, TV represents the total target volume, and VPI represents the volume receiving the prescription dose. Ideally, the CI should be close to 1. The D5% refers to the minimum dose received by 5% of the planning target volume (PTV) according to the DVH, indicating the maximum dose. Conversely, the D95% represents the minimum dose received by 95% of the PTV, indicating the minimum dose. A lower HI indicates better homogeneity.

Plan Verification
The MatriXX Universal Detector Array, manufactured by IBA in Germany, was utilized to compare the plan quality of all the plans in this study. The evaluation of plan quality involved calculating the gamma index and the gamma pass rate (GPR) by comparing the dose fluence generated by the TPS with the measurements obtained from the MatriXX detector.

The GPR was determined by dividing the number of measurement points that met the pre-defined criteria by the total number of measurement points within the specified threshold. To eliminate low-dose signals, a lower limit of 10% was set during the gamma calculation. The analysis of measurements was conducted using a 3% dose difference (DD) and a 3-mm distance to agreement (DTA).

In this study, the global gamma indices were considered clinically acceptable if the GPR for the 3%/3 mm criteria was equal to or greater than 95%. To facilitate the measurements, the immatrix detector was inserted into the miniphantom and positioned on the treatment couch. The iso-plane was set at the depth of the effective point of measurement on the side of MatriXX using the corresponding markers. MatriXX was calibrated at the used photon energy and all plan intended by calibration factors.

Statistical Analysis
To compare dosimetric parameters and measurement results between different groups, the Wilcoxon signedrank test was employed. The statistical analysis was performed using SPSS 22.0 software. A p<0.05 was considered statistically significant, indicating a notable difference between the groups.

Results

Target Doses
Figure 1 shows the DVH graph of VMAT plans with different MSW parameters for a typical patient. In MSW2.0 cases, PTV (primary) and PTV (LN) from the DVH failed to meet the clinical requirement, i.e., for PTV primary D95% = 90%, and for PTV lymph nodes, D95% = 91.2% (<95%).

Fig. 1: The dose volume histograms of four different VMAT plan with different MSW"s for a typical prostate cancer.
VMAT: Volumetric-modulated arc therapy; MSW: Minimum segment width.

Table 2 shows the comparison of the mean dose, maximum dose, HI, and CI of the target PTVs. The dosimetric parameters of PTVs" mean dose and maximum dose were comparable among four MSWs plans, but conformity and homogeneity were poor as MSWs value increased from 0.5 to 2.0. Therefore, all MSWs group plans were not comparable in terms of HI and CI. On average, a 15% decrement was found in conformity with MSW2.0 plans as compared to MSW0.5.

Table 2: Dosimetric results of PTVs for prostate VMAT plans with different MSWs (n=12)

As shown in Figure 2, VMAT plans with MSW of 0.5 cm and 1.0 cm were similar for all dosimetric parameters of PTVP and PTVLN (p>0.05) except for maximum dose to PTVP. Except for maximum doses to both targets, VMAT plans with MSW of 0.5cm were better as compared to VMAT plans with MSW of 1.5 cm and 2.0 cm (p<0.05). However, the maximum dose to the PTV (Primary) in the MSW0.5 plan was 0.9Gy higher than in the MSW1.0 (p=0.005).

Fig. 2: Graphical Representation of homogeneity index, conformity index, and mean dose with different MSW"s for both PTV"s for n=12.
PTV P: Primary planning target volume; PTV LN: Lymph node planning target volume.

OAR Doses
Figure 3 compares V58Gy(%), V54Gy(%), V50Gy(%), and V41Gy(%) doses to the bladder, rectum, and V35Gy(%) of the femoral heads among the four MSW groups. As MSW"s value increased, OAR doses decreased. There were no statistical differences in the OAR doses between the MSW1.0 and MSW0.5 plans (p>0.05). For OAR, there was no significant statistical difference in the OAR doses between the MSW1.5 and MSW0.5 plans except V58Gy for rectum and V41Gy for bladder as shown in Table 3. Except for bladder doses in higher MSW plans, there were no significant differences in doses between the four types of VMAT plans in terms of other remaining OAR.

Fig. 3: Graphical Representation of OAR"s doses with different MSW"s for n=12. OARs: Organ at risks.

Table 3: Dosimetric results of OARs for prostate VMAT plans with different MSWs (n=12)

MU and CP
As the MSW value increased, CP in the prostate cancer of the VMAT plan decreased; the mean number of CP for the plans with MSWs of 0.5 cm, 1.0 cm, 1.5 cm, and 2.0 cm was 274, 248, 204, and 188, respectively (Fig. 4). Moreover, the MUs of the VMAT plan decreased as the MSW increased (Fig. 4). The mean MUs for the plans with MSWs of 0.5, 1.0, 1.5, and 2.0 cm were 1503.7±80.4, 1102.4±76.7, 914.8±64.2, and 900.9±81.6, respectively.

Fig. 4: Graphical representation monitor unit, control points, GPR, and PDT with VMAT plan of different MSW"s for n=12.
PDT: Plan delivery time; GPR: Gamma passing rate.

Dosimetric Verification and Plan Delivery Time
The evaluation involved a comparison between the measured planar dose and the dose calculated by the TPS, employing the gamma passing criteria with a 3% DD and a 3 mm DTA. Table 4 illustrates the GPRs for plans characterized by MSW of 0.5 cm, 1.0 cm, 1.5 cm, and 2.0 cm. The GPR was most pronounced in the plan using 1.5 cm MSW, while it was least pronounced in the plan using 0.5 cm MSW. In addition, Table 4 shows the PDT for the 12 patients from the time the beam is turned on to the time it is turned off. CP and MUs of the VMAT plan decreased as MSW increased, as did plan delivery time.

Table 4: Statistical Results MU, Control Points, PDT, and GPR of VMAT plans with different MSW values (n=12)

Discussion

The process of designing VMAT plans to treat prostate cancer results in a significant number of long, small, and irregular segments.[14,15] The MSW plays a crucial role in optimizing the formation of these apertures. Unfortunately, these segments can sometimes lead to challenges during clinical delivery, causing low verification rates and even interruptions in the VMAT plan delivery.[9,16] To address the issue of plan complexity, we investigated the impact of varying MSW values on the quality of prostate cancer VMAT plans. We compared four different optimization schemes, each based on a different MSW value. The evaluation of plan quality involved assessing several parameters, including the HI, CI, maximum, and mean doses to the planning target volume (PTV), as well as the dose-volume indices of organ at risk, MUs, and CP. Our findings revealed that VMAT plans generated with an MSW of 1.0 cm exhibited similar dose distributions to plans with MSWs of 0.5 cm (Fig. 2). However, we observed that plans with MSWs of 1.5 cm and 2.0 cm displayed slightly inferior quality, failing to meet the clinical requirements adequately (Table 2).

In addition, the number of CP and MUs decreased as the MSWs increased (Table 4). When compared to the plan using an MSW of 0.5 cm, the mean MU reductions in the plans using MSWs of 1.0, 1.5 cm, and 2.0 cm were 26.68%, 39.16%, and 40.0%, respectively, while the total CP was decreased by 9.2%, 25.5%, and 31.3%, respectively. Previous studies showed that decreasing the MUs for treatment delivery reduces the constraint factor of the leaves" trajectories, the complexity of intensity? modulated radiation therapy plans, and treatment time[17-19] Hence, as the MSW increases and VMAT plan complexity decreases, the therapeutic efficiency may improve as well. The average delivery times of the plans using MSWs of 1.0, 1.5, and 2.0 cm were decreased by 37.8, 49.8, and 49.8 s, respectively (a drop of approximately 12%, 15.8%, and 15.8%, respectively), compared to the plan with an MSW of 0.5 cm.

The measured and computed doses were assessed through a matrix detector, and all treatment plans demonstrated favorable GPRs. The average GPR >94% with a 3% DD and 3-mm DTA threshold indicates strong congruence between measured and calculated doses.[20] Enhanced agreement between measured and TPS-calculated doses was observed with a higher MSW. This outcome was anticipated due to the decrease in the number of small fields as MSW increased, facilitating dosimetric verification.

Conclusion

We concluded that VMAT plans for prostate cancer generated with an MSW of 1.0 cm demonstrated comparable dose distributions to plans with MSWs of 0.5 cm. However, plans with larger MSWs showed a decline in quality, raising concerns about their clinical suitability.

Peer-review: Externally peer-reviewed.

Conflict of Interest: All authors declared no conflict of interest.

Ethics Committee Approval: The study was approved by the Amity University Ethics Committee (date: 02/11/2023).

Financial Support: None declared.

Authorship contributions: Concept - N.J., A.K., As.K.; Design - N.J., A.K., As.K.; Supervision - N.J., A.K., As.K.; Funding - N.J., As.K.; Data collection and/or processing ? N.J.; Data analysis and/or interpretation - N.J., A.K.; Literature search - N.J.; Writing - N.J., A.K.; Critical review - A.K.

References

1) Xia C, Dong X, Li H, Cao M, Sun D, He S, et al. Cancer statistics in China and United States, 2022: Profiles, trends, and determinants. Chinese Med J 2022;135(5):584-90.

2) Gay HA, Michalski JM. Radiation therapy for prostate cancer. Mo Med 2018;115(2):146?50.

3) Murgi? J, Fröbe A, Kiang Chua ML. Recent advances in radiotherapy modalities for prostate cancer. Acta Clin Croat 2022;61(Suppl 3):57-64.

4) Teoh M, Clark CH, Wood K, Whitaker S, Nisbet A. Volumetric modulated arc therapy: A review of current literature and clinical use in practice. Br J Radiol 2011;84(1007):967-96.

5) Rashid A, Ahmad Z, Memon MA, Hashim ASM. Volumetric Modulated Arc Therapy (VMAT): A modern radiotherapy technique - A single institutional experience. Pak J Med Sci 2021;37(2):355-61.

6) Wang Y, Chen L, Zhu F, Guo W, Zhang D, Sun W. A study of minimum segment width parameter on VMAT plan quality, delivery accuracy, and efficiency for cervical cancer using Monaco TPS. J Appl Clin Med Phys 2018;19(5):609-15.

7) Hong J, Han JH, Luo HL, Song YQ. Optimization of minimum segment width parameter in the intensitymodulated radiotherapy plan for esophageal cancer. Int J Gen Med 2021;14:9913-21.

8) Nithiyanantham K, Kadirampatti Mani G, Subramani V, Karukkupalayam Palaniappan K, Uthiran M, Vellengiri S, et al. Influence of segment width on plan quality for volumetric modulated arc based stereotactic body radiotherapy. Rep Pract Oncol Radiother 2014;19(5):287-95.

9) Moon Y, Bae S, Choi C, Jeon W, Kim J, Lee M, et al. Effect of minimum segment width on gamma passing rate considering MLC position error for volumetric modulated arc therapy. J Korean Phys Soc 2019;74:724-30.

10) Peng Z, Aowen D, Mingying G, Liu Y, Li J, Luo J, et al. The effect of minimum segment width of collimator on VMAT plan quality for rectal cancer using Monaco TPS. Available at: https://www.researchsquare.com/ article/rs-1632761/v1. Accessed Nov 15, 2023.

11) Sun H, Wang N, Wang X, Huang G, Chang Y, Liu Y. A study of different minimum segment area parameters on automatic IMRT plans for cervical cancer using Pinnacle3 9.10 TPS. Medicine Baltimore 2022;101(36):e29290.

12) Yoosuf AM, Ahmad MB, AlShehri S, Alhadab A, Alqathami M. Investigation of optimum minimum segment width on VMAT plan quality and deliverability: A comprehensive dosimetric and clinical evaluation using DVH analysis. J Appl Clin Med Phys 2021;22(11):29-40.

13) Wu M, Jin J, Li Z, Kong F, He Y, Liu L, et al. Influence of beamlet width on dynamic IMRT plan quality in nasopharyngeal carcinoma. PeerJ 2022;10:e13748.

14) Nguyen M, Chan GH. Quantified VMAT plan complexity in relation to measurement-based quality assurance results. J Appl Clin Med Phys 2020;21(11):132-40.

15) Chiavassa S, Bessieres I, Edouard M, Mathot M, Moignier A. Complexity metrics for IMRT and VMAT plans: A review of current literature and applications. Br J Radiol 2019;92(1102):20190270.

16) Moon YM, Bae SI, Han MJ, Jeon W, Yu T, Choi CW, et al. Correlation between average segment width and gamma passing rate as a function of MLC position error in volumetric modulated arc therapy. Technol Cancer Res Treat 2021;20:15330338211059937.

17) Reddy BR, Varatharaj C, Ravikumar M, Ganesh KM. Effect of fluence smoothening in intensity modulated radiotherapy planning and delivery. Int J Med Res Rev 2018;6(6):292-300.

18) Sun X, Xia P, Yu N. Effects of the intensity levels and beam map resolutions on static IMRT plans. Med Phys 2004;31(9):2402-11.

19) Athiyaman M, Hemalatha A, Chougule A, Joan M, Kumar H. Estimation of monitor unit through analytical method for dynamic IMRT using control points as an effective parameter. J Radiother Pract 2022;21(3):343-350.

20) Madhusudhana Sresty NVN, Raju AK, Reddy BN, Sahithya VC, Mohmd Y, Kumar GD, et al. Evaluation and validation of IBA I"MatriXX array for patient-specific quality assurance of TomoTherapy®. J Med Phys 2019;44(3):222-7.